1. Temperature up to 1400 degree
2. Heating speed up to 50 degree per min.
3. Microwave heating.
4. No pollution to material
HAMiLab-T system
High temperature microwave sintering furnace
Description
HAMiLab-T is our 3rd generation high temperature microwave laboratory system. It's eloquent and sophisticated design accents this system's advanced state-of-the-art technology. It is composed of 4 sub-systems that are integrated to create a truly hands-off experience of processing materials and metals, chemical and biological samples. The only standard in manufactured high-temperature, high-power research furnaces. HAMiLab-T's stainless steel heating chamber is attached to vacuum and gas-infiltration systems that provide a highly controlled environment for processing samples. Energy is delivered to the sample loads by direct energy transfer form a high precision micro head, capable of delivering up to 6kw of microwave power to heat samples to temperatures only limited by the state-of-the-art in insulation materials.
Application field:
- Specially apply for synthesis of high-purity metal nitride
- Catalytic decomposition and reduction on SO2 and NOX
HAMiLab-T can also be used in following fields:
1. Compounding and sintering of inorganic powder
1). Carbide: SiC, CrC, VC, etc
2). Nitride: Si3N4, MnxNy, AlN, VN, CrN, etc
3). Electronic ceramic powder: barium titanate, barium-strontium titanate, strontium titanate, barium zirconate titanate, etc.
2. Inorganic material/products sintering
3. Incinerating
4. Melting and thermal treatment
5. Melt oxidized ore carbon thermal reduction
Technical specification
Technical specification |
Model |
HAMiLab-T3000 |
Voltage |
AC 380 plus/minus 10V 3Ø, 50/60Hz, 20 Amps |
Rating power |
6KW |
Microwave output power |
0.3-2.85KW continuously variable |
Microwave frequency |
2.45GHzplus/minus 25MHz |
Max working temperature |
1400 Celsius degree |
Temperature measuring way |
IR thermometer (U.S.A. Raytek) |
Range of temperature measuring |
450 Celsius degree -2250 Celsius degree. |
Temperature accuracy |
plus/minus 0.5% |
Static vacuum degree |
less than 100Pa |
Atmosphere system |
air, oxygen, nitrogen, argon, weakly reducing atmospheres, etc |
Oven size |
Ø380*520mm |
Max. loading space |
80 x 80 x 60mm |
Cycle cooling water flow |
more than 1.5m3/h |
Cycle cooling water pressure |
more than 0.15MPa |
Cycle cooling water total hardness |
less than 60mg/L |
Control system |
embedded system and touch screen, connected with PC, with data memory and printing function |
Microwave leakage |
less than 1mw/cm2, (national standard is less than 5 mw/cm2) |
Floor area |
1.5 m3 |